Specification
規格 | 斜度 | L | l1 | D | 螺桿尺寸 | 重量 |
|---|---|---|---|---|---|---|
| CHW-SK30 | NT30/BT30 | 650 | 500 | 210 | M12x1.75P | 50.00 |
| CHW-SK40 | NT40/BT40 | 680 | 500 | 240 | M16x2.0P | 70.00 |
| CHW-SK50 | NT50/BT50 | 900 | 700 | 260 | M24x2.0P | 110.00 |
Operating Instructions
▍Operation Precautions
- Check if the calibration bar is standard and ensure the rotating spindle is clean.
- Locate the fixing key on the spindle and secure the rotating shaft to the right.
- Lock and calibrate the screw (approx. 15mm thread depth) and tighten the fixing nut securely.
- Flip the spindle fixing key to the left to allow the spindle to rotate.
- Place the dial indicator at the inspection position to perform inspection and calibration.
- During measurement with the dial indicator, ensure the highest point of the test bar is captured as the probe contact point.
▍Precautions
- The calibration base must be placed on a horizontal fixing mount.
- Keep the storage area clean and maintain room temperature at around 20° ±10°.
- Strictly prohibit striking or touching the calibration base.
- Maintain cleanliness at all times and perform regular inspections and maintenance.
- Place the dial indicator at the inspection position to perform inspection and calibration.
- During measurement with the dial indicator, ensure the highest point of the test bar is captured as the probe contact point.